HeartStart Intrepid Instructions for Use Addendum
English
This document is an addendum to the HeartStart Intrepid Instructions for Use. This addendum
provides the following information:
•
Updated information related to AC power, REACH regulation, and Proposition 65.
•
Installing the Intrepid docking station on an ambulance shelf.
•
Attaching and removing the Intrepid from the docking station.
•
Cleaning instructions for the docking station.
•
HeartStart Intrepid weight specification.
•
Prehospital power source warning.
Keep this addendum with your Instructions for Use.
Safety Consideration
Use only a 3-wire AC power cord with a 3-pin grounded plug.
CAUTION: The HeartStart Intrepid must be plugged into a hospital-graded outlet with a protective ground.
Failure to do so may cause electric shock.
Symbol Definitions
The Class II equipment symbol is applicable when the HeartStart Intrepid is connected to DC
power only.
Figure 1
AC Power Specifications
Power: –
AC Line Powered: 100 – 240 VAC, 50 or 60 Hz, 1.8 – 0.75 A, Class I equipment
Table 1
Surge line to ground IEC 61000-4-5
Class II Equipment Symbol
AC power input from grounded electric outlet.
–
Rechargeable Lithium Ion battery.
Input AC Power Ports
Immunity Test
HeartStart Intrepid Instructions for Use Addendum
Immunity
Compliance
Test Level
Level
±0.5 kV,
±0.5 kV,
±1 kV,±2kV
±1 kV,±2 kV
Electromagnetic
Environment Guidance
Emergency medical services
environment
Professional healthcare facility
environment
1