All external parts of the hardware should be disinfected with 70% isopropyl
alcohol wipes. Under normal conditions, it is not necessary to remove
CORE™ attachment from the stethoscope tubing during the disinfecting
procedure.
NOTE: DO NOT immerse the device in any liquid or subject it to any high-
pressure/autoclave sterilization processes.
It might be necessary to remove the Eko CORE™ Digital attachment from
the Stethoscope, if the device is excessively soiled or a fluid invasion is
suspected. The following disassembly procedure is recommended:,
1. Pull the stethoscope tubing off of the metal stem of the CORE™
attachment on both ends.
2. Wipe all parts of the stethoscope clean with 70% isopropyl alcohol wipes
or disposable hospital-grade disinfectant wipe, or disposable wipes with
soap and water including CORE™'s surface, stethoscope tubing, tubing
connector, eartips and chest piece.
3. A 2% bleach solution may be used to disinfect your stethoscope tubing,
tubing connector, and chest piece; however, the tubing may become
discolored after exposure to bleach.
4. To prevent staining of stethoscope tubing, avoid contact with pens,
markers, newsprint, or other printed material. It is good practice to wear
your stethoscope over a collar whenever possible.
5. Reassemble the stethoscope by reinserting the metal stems of the CORE™
attachment into the stethoscope tubing as described above in the
installation section.
11. Operating Conditions
Environmental conditions of transport and storage between uses -40° to
+55°C, relative humidity range of 15% to 93% at an atmospheric pressure
range of 700 hPA to 1060 hPa (conforming to IEC 60601-1-11).
Continuous operating conditions
-30° to 40°C; relative humidity range of 15% to 93%, at an atmospheric
pressure range of 700 hPa to 1060 hPa (conforming to IEC 60601-1-11)
Avoid exposure to extreme heat, cold, solvents and oils. Extreme heats and
colds will negatively affect the lithium ion battery in the device and may
affect battery life.
No Modifications
Failure to follow care and maintenance recommendations could result
in damage to the internal components of CORE™. Internal damage to
the product could cause malfunction of the product, which may lead to
complete loss of function. If problems are encountered with CORE™, do not
attempt to repair it. Please notify our support team for assistance.