Fault indication by LEDs
LED
.
Display
.
Cause of fault
To eliminate fault
Diagnostic
Normal
If fault
Desired
is given
value
212V
"Perhaps
#5 - 104
Voltage controller
Replace
red
or microprocessor
pP circuit board
(Without
circuit board
ing)
+5V UE
Dark
red
5\ #5%
1. Voltage controller
Change voltage controller
Cable 87 or
ST7
circuit board
2. Fuse F3
Change fuse on analogue
3. Short circuit on the
circuit board
analogue or micro-
Change analogue or micro-
processor circuit
processor circuit board
4. Flat cable 803, 8012,
$13
+5V ST
Dark
Red
6V 25%
1. Voltage controller
Change voltage controller
cable BU?
circuit board
2. Fuse F4
Change analogue or micro-
3. Short cirovit m the
processor circuit board
analogue or micro-
processor circuit
board
4. Flat cable BU3, BUIZ
*6V
Green
Red
6V £5%
1. Voltage controller
Change voltage controller
cable BU7
Circuit board
2. Analogue circuit
board
3. Flat cable BU3, BUIZ
6 2 UE
Flashes
Does not
983,04 KHz
Change microprocessor
every
flash
circuit board
second
9 2
Flashes
Does not
1228,8 KHz
Microprocessor
Change
every
flash
circuit board
second
MF
Flashing |
Does not
According to
Microprocessor
Change
frequency
flash or
delivery rate
circuit
board
dependent | stays lit
original MF
of the pP cir-
arit board
streched over
a monofiop
Drese Serwce-Unteriage sarf ohne prsere Genenmaguno
.
Seite
Aus
abe
μαι
aca werden ns Diept Osten Meares
Service-Manual
Page
rd
B Braun Melsuncen
neony
TECHNEMeEC
Ko
big
nan
INF.secura 999ml/h
12
03/89
seniz
>
i
= CONSENT, ang FEMS the property di oor company.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis