Faults and fault clearance
Alarm
Fault
Remedy
red LED activated
audible alarm
delivery rate display shows
AAA
Volume display shows
F4F4
flow sensor positioned
still on handle
switch off unit
place flow sensor on drip
chamber
switch on unit
flow sensor disconnected
switch off unit
connect flow sensor
switch on unit
Flow sensor defective
switch off unit
change flow sensor
switch on unit
. flashing battery indicator,
i audible alarm
low battery capacity
(capacity still appr. 10 min.)
connect unit to mains
all displays and indicators
not illuminated
no capacity
disregarded low-capacity-
alarm
connect unit to mains.
unit should remain
connected to mains for
the next 20 hours
audible alarm
delivery rate display
' shows delivery rate and
: 001 alternating
| volume display flashes
preselected volume
KVO-operation
preselected volume has
been infused.
The INFUSOMAT? secura
is operating on KVO
(1 ml/h) for max. 30 min.
audible alarm
delivery rate display shows
delivery rate and 000
' alternating
end of KVO-operation
INFUSOMAT secura has
stopped after 30 minutes
of KVO-operation.
Additional alarm is given
on the staff call system,
| volume display shows
if connected
preselected volume and
EEEE alternating
undefined displays
malfunction of unit
call for service
The audible alarm can be muted for 2 min.
by pushing the key
31