Faults and fault clearance
Alarm
Fault
Remedy
red LED activated
audible alarm
Delivery rate display shows
A 40
infusion container empty
replace infusion container
restart unit
roller clamp closed
open roller clamp
restart unit
occlusion
a. check infusion line
for bends.
b. check position of
cannulae/catheter
c. check cannulae/catheter
for blockages
restart unit
condensed drip chamber
eliminate condensation by
shaking drip chamber
restart unit
Flow sensor tilted
Fix drip chamber in vertical
position
restart unit
free-flow
Close roller clamp,
check correct position of
infusion line,
if necessary use new
infusion line.
Open roller clamp (no
continous driping allowed)
restart unit
red LED activated
audible alarm
Delivery rate display shows
A 50
Air in Line
a. Deaerate infusion line
(if necessary use other
infusion line)
b. Set fluid level in drip
chamber
restart unit
incorrect position of infusion
line
check position of infusion
line according to instructions
restart unit
incorrect closed pump door
close pump door
restart unit
The audible alarm can be muted for 2 min.
by pushiung the key
30