Assembly
Optional accessories
Safety hood
Caution:
Possible damage of the protective material
At to high temperatures and/or unsuitable heating bath liquids the
material becomes soft and may deform and possibly crack.
The safety hood is only suitable for heating bath temperatures < 130° C.
(For suitable heating bath liquids please refer to Chapter "Start Up,
Heating bath and flask lift" and chapter "Accessories and Spare parts,
Accessories").
Before you can use the safety hood, you need to screw the support of the safety hood to
the height stop.
➜ Unscrew the locking screw from the
height stop band.
➜ Slide the safety hood off the support.
➜ Tighten the safety hood carrier on
the height stop band using the two
supplied Allen keys.
✓ The knurled screw on the safety hood
carrier takes over the function of the
locking screw for the height stop band.
➜ Loosen the clamping screw and slide
the safety hood horizontally onto the
carrier until the hood sits above the
heating bath.
➜ Tighten the clamping screw.
✓ The safety hood is installed. The safety
hood can be lifted open at the handle
until it clicks into place.
Safety hood carrier
Allen screw
Safety hood
carrier
97
Height stop
Knurled screw
Clamping screw