Assembly
Connecting peripheral devices

Connecting cooling water
Assembly parts:
▪ 2x tube nipple
(included in scope of delivery)
▪ 1x tube, Ø inside 7-8 mm
(not included in scope of delivery)
Coolant inlet and outlet are located on the
back and bottom of the condenser.
➜ Remove the screw caps from the
cooling water inlet and outlet.
➜ Push the screw caps over the tube
nipples so that the black seal is located
inside the cap.
➜ Push the respective inlet and outlet
hoses for the cooling water through
the screw cap over a tube nipple.
Tighten the screw caps with the seal
first on one of the respective openings
of the condenser.
✓ Cooling water inlet and return line are
connected.
Warning:
Risk of explosion
A closed valve in the recirculation system will build up high pressure
in the glass coils and may result in glass bursting.
The cooling coils of the condenser may be exposed to a maximum
cooling water pressure of 2 bar. To prevent a higher pressure from
ever occurring, no valve or throttle must be installed in the cooling
water return line.
Tube nipple
Cooling water inlet and outlet
94