Troubleshooting
Error messages
1
2
3
5
Resetting temperature limiter

➜ Switch the device off Remove the
evaporating flask
➜ Disconnect the heating bath plug
remove the heating bath.
➜ If necessary, empty the heating bath
and turn it over.
➜ Use a suitable object to briefly press
the red button on the bottom until it
clicks into place.
✓ Reset the temperature limiter
If you experience a malfunction that cannot be resolved using the information listed
above, please inform your authorized Heidolph dealer immediately.
Possible cause
Only for motor lift ver-
sion:
The transportation lock is
still activated.
The cable connection
between the control panel
and the base device gas
bee separated.
▪ Heating defective
▪ Overheat protection
activated
Overtemperature
protection
+5 °C exceeded
Heating bath fluid
completely evaporated
Heating bath sensor
difference
> 10 °C
▪ Temperature sensor
heating bath defective
▪ Temperature of the
heating bath fluid > 217
°C
▪ Temperature of the
heating bath fluid < 0 °C
Motor overloaded
Remedy
▪ Remove transportation lock
▪ Move lift upward
▪ Switch the device off and then on
again
▪ Re-establish the cable connection
▪ Switch the device off and then on
again
▪ If necessary, allow the heating bath
to cool down and empty, reset the
overheat protection (see "Resetting
temperature limiter" section).
Contact local dealer
▪ Switch the device off
▪ Allow the heating bath to cool
down, reset the overheat protection
(see "Resetting temperature
limiter" section).
▪ Filling with heat transfer liquid
Contact local dealer
Contact local dealer
▪ Switching the device off
▪ Contact local dealer
Temperature limiter
80

Werbung