Owner's Guide
- Ensure that the power adapter is plugged into the
charger base securely. Ensure that the AC power cord
is plugged into the power adapter and the wall outlet
securely.
If the Shot or Mode button lights up but doesn't
work properly
- The curing light may be discharged and needs
recharging.
- Place the handpiece on the charger base to recharge
until the red light changes to the green illuminated
light indicating a fully charged curing light.
- If the curing light does not activate, the unit requires
service. Please contact your supplier or authorized
service center.
If the S.P.E.C. 3
®
LED Curing Light doesn't
polymerize light-cured materials well
- Ensure that a blue LED light is emitted from the light
guide. Do not look directly into the light output.
- Make sure the light guide is inserted into the opening
on the handpiece completely. Even a slight gap
between the light guide and LED may reduce the
polymerization light output up to 50%.
- Check the light guide for any damage. If the light
guide has an inner crack even though its surface looks
good, it may interrupt proper light emission.
- Check if there's any debris, such as resin residue or
sealant on the tip of the light guide.
- Ensure that appropriate mode and time setting is
selected according to the type of light-cured material
being used.
- Make sure the light cured dental material is stored
according to the manufacturer's suggested storage
and that the date on the material's package has not
expired.
EN
- Check curing depth and time using the enclosed test
block. See Composite Curing Test Block Instructions.
(Section 5.5)
If the LED light is flickering or unstable
- Please contact our customer service representative for
assistance.
VII. Cleaning, Disinfecting and Sterilizing
Cleaning and Disinfection Instructions must be followed
after each use. Polyethylene film barrier sleeves supplied with
the S.P.E.C. 3
®
LED Curing Light are single use and non-sterile.
The purpose is to aid in infection control by addressing cross
contamination. Ensure that a new undamaged barrier sleeve is
installed each time the S.P.E.C. 3
Suggested surface disinfectant:
CaviCide
®*
(or equivalent quaternary/alcohol blend
disinfectant)
Do NOT use:
Glutaraldehyde
Denatured alcohol
Lysol
®*
Phenol or phenolic cleaner
Ammonia complex
Iodine complex solutions
Disassembly and Inspection:
1. Remove the red elastomer rubber light shield or eye shield
from glass fiber-optic light guide. Examine for damage.
2. Remove and discard used polyethylene barrier shield.
®
LED Curing Light is used.