XI. Electromagnetic Compatibility
The following are guidance and declarations regarding electromagnetic compatibility for the S.P.E.C. 3
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The S.P.E.C. 3
®
LED Curing Light is intended for use in the electromagnetic environment specified below. The customer or the end user of the
S.P.E.C. 3
®
LED Curing Light should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11:2004
RF emissions CISPR 11:2004
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/Flicker emissions IEC
61000-3-3
EN
11.1 EN/IEC 60601-1-2:2007 Table 1
Compliance
Group 1
Class B
Class A
Complies
®
LED Curing Light.
Electromagnetic environment - guidance
The S.P.E.C. 3
®
LED Curing Light uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
®
The S.P.E.C. 3
LED Curing Light is suitable for use in all
establishments other than domestic, and may be used in
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
for domestic purposes, provided the following warning is
heeded:
Warning: This equipment is intended for use by healthcare
professionals only. This equipment may cause radio interference
or may disrupt the operation of nearby equipment. It may be
necessary to take mitigation measures such as re-orienting or
relocating the S.P.E.C. 3
®
LED Curing Light or shielding the
location.
S.P.E.C. 3
®