6
Take caution when using this product in the
presence of a flammable anesthetic mixture or an
oxidizer like oxygen or nitrous oxide. This device
may act as an ignition source.
DO NOT expose soft oral tissues at close proxim-
ity or burns may occur. Maintain a safe distance
between the light guide tip and the soft tissue.
If using the S.P.E.C. 3
Standard Mode and in close proximity of the gin-
gival tissue, DO NOT expose tissue for more than
5 seconds or minor burns may occur. If longer
curing time is required, consider a dual-cure prod-
uct (composite or adhesive). Irradiance output in
High Power Mode is 1600 mW/cm².
In 3K Mode, DO NOT expose soft tissue for more
than 2 seconds or burns may occur. If a longer
cure is needed, consider a dual cure product
WARNING
(composite or adhesive). Irradiance output in 3K
mode is 3000mW/cm
Barrier sleeves are single patient use only. Barrier
sleeve must be used to prevent patient cross
contamination.
Persons having a history of photosensitizing
drugs should not be exposed to light from this
light.
Individuals with a history of cataract surgery
may be particularly sensitive to the exposure
of light and should be discouraged from
treatment unless adequate safety measures
are undertaken. Individuals with a history of
retinal disease should seek advice from their
ophthalmologist prior to operating the device.
Adjust curing techniques in accordance with the
increased energy. Pulpal and soft tissue damage
may occur if excessive energy is applied to a
restoration.
Failure to comply with these warnings may
result in injury to personnel.
®
LED Curing Light in the
2.
EN
2.5. Safety Notes
1. Before use, check if the device is working properly and has
no visible damage.
2. The fiber-optic light guide is provided in a non-sterile
condition and must be sterilized prior to patient contact.
3. In case of device malfunction, discontinue use and
refer to the 'Trouble shooting' section in this Owner's
Guide. If the problem persists, contact our service center
immediately.
4. Do not attempt to repair, disassemble or remodel
this product without the manufacturer's permission
(COLTENE). Otherwise, product warranty will be voided.
5. The handpiece is not autoclavable. Do not sterilize the
handpiece by immersion or heat sterilization such as dry
heat, steam under pressure, or unsaturated chemical
vapor (fiber-optic light guide and light shield can be
autoclaved).
6. Using barrier sleeves will reduce the light output by
5-10%.
2.6 Environment Protection
This appliance is labeled in accordance with European directive
2002/96/EC concerning used electrical and electronic appliances
(waste electrical and electronic equipment – WEEE). This guideline
determines the framework for the return and recycling of used
appliances as applicable throughout the EU. The symbol on the
product, or on the documents accompanying the product,
indicates that this appliance may not be treated as household
waste. Instead it shall be handed over to the applicable collection
point for the recycling of electrical and electronic equipment.