CERTIFICATE OF CALIBRATION
The under-mentioned item has been calibrated at the following points in the Michell Instruments' Humidity Calibration Laboratory against
Test Equipment traceable to the NATIONAL PHYSICAL LABORATORY, Middlesex, United Kingdom and to the NATIONAL INSTITUTE OF
Certificate Number
Acknowledgement Number
Test Date
Test Equipment
Calibration Temperature (
Work Instruction Number
Reference (ppm)
0.130
0.291
1.401
5.371
17.51
58.26
111.4
. 204.7
616.3
Comments:
N/A
Calibration PASS. The results are within specification of the analyzer at the measured points detailed.
Traceability to National Physical Laboratory is over the range -90
Traceability to National Institute of Standards and Technology is over the range -75
Approved Signatory:
Michell Instruments
QMA 601
STANDARDS & TECHNOLOGY, Gaithersburg, Maryland, USA.
0
0
Q0433
°
C)
21 ( +/-2)
520
Abbildung 45
Typisches QMA601 Kalibrierzertifikat
Analyzer Serial Number
Oven Serial Number
Beat Frequency (Hz)
MG Moisture Level (ppm)
Inlet/Sensor Pressure (barg)
Outlet/Back Pressure (barg)
Measured Moisture
Content (ppm)
0.128
0.289
1.469
5.728
18.62
59.96
108.7
198.6
591.6
°
C to +90
°
C. (0.095 to 2253559 ppmV)
°
°
C to +20
C. (1.204 to 23632 ppmV)
Date of Issue: