Herunterladen Inhalt Inhalt Diese Seite drucken

Wire Guide And Fastening - MICRO-EPSILON wireSENSOR WDS series Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Installation and Mounting
4.3
If the measuring wire has to be extracted from the sensor to
A measuring wire
guide the wire respectively to fix it to the target,
under tension
- the sensor may not be held by another person
where operators are
- the measuring wire may not be further extracted but only
standing can lead to
injuries.
- the surroundings of the sensor have to be protected
> Danger of dam-
age to wire and
sensor.
Do not twist the
measuring wire.
Misalignment is only permissible up to 3 degrees.
Dragging of the measuring wire on the inlet hole or other
objects leads to damage and/or snapping of the measuring
wire.
i
wireSENSOR, WDS P60/P96/P115

Wire Guide and Fastening

to the specified measuring range
against snapping of the measuring wire
Fix the measuring wire to the target using a wire clip.
Guide the measuring wire vertically out of the sensor
housing.
If the measuring wire cannot be fed vertically out
of the housing, it is essential to use a guide pulley
(accessory TR1-WDS and TR2-WDS, see Chap.
Keep the measuring wire in an area where it cannot be
snagged or otherwise be violated.
max. 3 °
A
1).
Fig. 17 Wire fastening and misalignment
Wrong
Correct
Wire outlet 0 °
±3 ° tolerancy
Page 80

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis