Error messages and malfunctions

Messages appear in the display for several errors and malfunctions; causes and corrective
measures for these and other faults are listed below.
Error message
Prism not clean
No sample
Busy !
Protected !
Out of range
Temp Error
Lamp NG
Malfunction
I
nstrument can not
n
be switched o
Stored data
have been lost
(e.g. date)
Measurements of
same sample not
reproducible
Blank display
RA-510M
METTLER TOLEDO
Cause
You have not cleaned the prism
for a zero point calibration
You have no sample in the
measuring cell and have
pressed the Meas. key
Key is locked! You have, e.g.
pressed the Method key during
a measurement
The method is protected! You
have pressed the Meas. Para. ,
Calcu. Para. or Print Para. key
Value below or above measure-
ment range for this sample
The temperature sensor of the
thermostat control is faulty
The display is faulty
Possible cause
Cable not plugged in properly,
fuse blown
Internal battery dead
Prism faulty
Display contrast is set to mini-
mum or maximum
Measure
First clean the prism and then
calibrate
First add the sample
(Press Reset , if you wish to
abort the measurement)
Remove protection as appropri-
ate (select Method , see Section
7.2)
Switch the instrument off and
call
METTLER TOLEDO
Switch the instrument off and
call
METTLER TOLEDO
Measure
Check power cable, on/off switch
and fuses
Switch the instrument off and
call
METTLER TOLEDO
Switch the instrument off and
call
METTLER TOLEDO
Change contrast setting (see
Section 4)
Error messages
service
service
service
service
35