Quick Setup Guide TDL GPro™ 500

5 Error messages

Message
No sensor on channel 3
Signal Processing Failed
Laser Source Error
Bad Signal Quality
Flashcard Error
Pressure Input Error
Temperature Input Error
Configuration Mode
The GPro™ 500 error messages can be found in the M400 under the following path:
Menu ➝ Service ➝ Diagnostics ➝ TDL ➝ Messages
26
Comment
The M400 is unable to detect any of the ISM sensor(s) it can identify.
If no sensor is found it will disply the message NO SENSOR DETECTED
Fitting of the line profiles failed.
The laser wavelength has shifted. Readjustment of the laser tempera-
ture necessary
Transmission lower than 5% threshold
Missing or bad calibration and / or database data
Pressure reading out of extended range: 0.6 bara < P < 8 bara
4 – 20 mA input error: 4 mA > P > 20 mA
Pressure reading out of extended range: –20°C<T<1000°C
4 – 20 mA input error: 4 mA > P > 20 mA
Ethernet port in use: diagnostic or configuration in progress