6.3

Glucose solution (example)

– Prepare a 10% glucose solution.
– Add approx. 0.2 mL to the prism using a polyethylene pipette and stir carefully; the sample
must cover the entire surface of the prism. The refractive index is shown, the temperature
has not yet been reached.
– Press the Meas. key:
-MEASUREMENT-
nD
-RESULT-
nD
Report (GA42 printer) of the result for the first sample of a series (see Section 7.3.4)
Model:RA-510M
S/N
:UDB26543
Sample:
Name:Ch. Iserland
No.
01 - 01
Temp
:
20.00 C
nD
: 1.34768
Notes
a: Each result is stored and can be changed with the "Recalculation" function, printed out
again with the "Statistics" function and viewed with the "Delete data" function (see Sections
8.2, 8.3 and 8.5).
b: The Calib. key is locked if you measure a sample, the Meas. key is locked when you
perform a calibration. If you press the Calib. key or the Reset key during a calibration, this
will be aborted.
c: For solutions with a relatively high viscosity, we advise adding the sample using a plastic
spatula or spoon.
RA-510M
METTLER TOLEDO
[METHOD 0]
1.34771
°C
Temp.
19.
[METHOD 0]
1.34768
Temp. 19.97 °C
You may write the substance name onto this line
As long as the temperature of 20 °C
fluctuates by ± 0.1 °C, the value flashes.
The decimal places are not shown.
When the temperature is stable, the re-
fractive index is measured after approxi-
mately one minute (default value) and
the result is printed out.
Measurements