Functions
8.2
Recalculation (function 1: Re-Calcu.)
You can change or recalculate each result with this function by changing either the refractive
index of the measured sample or, for concentration results, the calculation function of the
active method in advance. The old values are overwritten, also for statistical evaluations.
-FUNCTION-
1.Re-Calculation
Low Samp. No: [01]
nD: [
– As soon as you have changed the nD value and confirmed this with ENTER, a new Brix or
Conc value is displayed. The Conc value is calculated with the current calculation para-
meters.
– If you change only the calculation function in advance, the recalculated Conc value is
displayed as soon as you have confirmed the sample number.
Note: If no result is available for the entered sample number, "[- - - - - - - -]" appears for the
nD value.
8.3
Automatic statistics evaluation (function 2: Auto Statis.)
The RA-510M evaluates the results of a series automatically and stores the statistical data
mean value, standard deviation and relative standard deviation.
Each method has its own result memory whose statistical values are cleared only after you
have entered a new series number for this method (see Section 7.3.4) or reset the
parameters to default (see Section 8.9). You can delete individual results of a series, e.g.
outliers which should not be included in the statistical evaluation (see Section 8.5).
Example of 8 measurements with nD as result
-FUNCTION-
2.Auto Statistics Calcu.
Data Print :
Unit
Mean
SD
CV
24
]
Off
On
:
8
: 1.33299
: 0.00001
:
0.0004 %
– Enter the number of the sample whose
result you wish to change and confirm
with ENTER:
The measured refractive index is dis-
played and, depending on the result,
the Brix value or Conc value.
Statistical data are printed out only when
you press the Print key.
Sample numbers with the corresponding
results and statistical data are printed
out.
Number of evaluated results
Mean value
Standard deviation
Relative standard deviation (coefficient
of variation)
METTLER TOLEDO