Error messages

9.
Error messages
Error 1 - Conductivity measuring value out of range
Check if the sensor is properly connected and placed in a sample solution.
Error 2 - Temperature measuring value out of range (–5...105 °C)
Keep the sample temperature within the range.
Error 3 - The measured calibration standard temperature is out of the range (0...35 °C )
Keep the calibration standard temperature within the range.
Error 4 - In nLF temperature correction mode, temperature measuring value out of range (0...35 °C)
Keep the sample temperature within the range.
Error 5 - Self-diagnosis failed
Repeat self-diagnosis and make sure that you press the correct keys while the
icon is
blinking. If Err 5 still appears, call METTLER TOLEDO service.
12
METTLER TOLEDO SevenEasy™ conductivity