5.5.4 Printing from Memory
When scrolling through the memory (see "Recalling from Memory (page 11)") you can print the entry
that is currently viewed by pressing and holding the STO/RCL key for 2 s.
The printout format is following:
Recall 01:
EP, Value, Unit, Temp.
ME 4.01pH 35.6ºC ATC
5.6 Self-Diagnosis
1
Press and hold Read and Cal simultaneously until the meter displays the full screen.
Each icon blinks one after the other. This way you may check whether all icons are correctly
shown. The next step is to check that the keys are functioning correctly. This requires user
interaction.
2
When
blinks, five icons are displayed. Press the five keys in any order.
Each time you press a key an icon disappears from the screen.
3
Continue to press the other keys until all the icons have disappeared.
When the self-diagnosis has been completed successfully, PAS appears. If self-diagnosis fails,
error message Err 1 appears (see "Error Messages (page 13)").
Note
You have to finish pressing all five keys within two minutes, otherwise Err 1 appears and you will have
to repeat the procedure.
5.7 Error Messages
Error 0
Memory access error
Error 1
Self-diagnosis failed
Error 2
Measured values out of
range
Error 3
Measured buffer temperature
out of range (5 ... 40 °C)
Error 4
Offset out of range
Error 5
Slope out of range
Error 6
Meter cannot recognize the
buffer
Reset to factory settings
Repeat the self-diagnosis procedure and make
sure that you finish pressing all five keys within
two minutes.
Make sure that the electrode wetting cap has been
removed and the electrode is properly connected
and placed in the sample solution.
If no electrode is connected, put the shorting plug
in the socket.
Keep the buffer temperature within the range for
calibration
Make sure you have the correct buffer and that it
is fresh.
Clean or replace the electrode.
Make sure you have the correct buffer and that it
is fresh.
Clean or replace the electrode.
Make sure you have the correct buffer and that it
is fresh.
Check that the buffer has not been used more
than once during the calibration.