Herunterladen Inhalt Inhalt Diese Seite drucken

Record Parameters - Mettler Toledo RA-510M Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7.3.3

Record parameters

– Press the Print Para. key (method 1 with result "Conc" is selected):
-PRINT PARA.-
Format:
Fix
nD Form (0-5)
Conc. Form (0-5):[3]
with 5 places after the point, Conc. with 3 places; if you confirm "Fix" with ENTER you can
change this (see Notes).
Off:
No result is printed out.
Variable: You can select the parameters which should be recorded if you confirm this setting
with ENTER:
-PRINT PARA.-
Sample:Off
On
Date
:Off On
Lot
:Off On
Method:Off On
Temp
:Off On
M.Time:Off On
• Series/sample number
• Date
• Lot number
• Method number
• Temperature
• Measurement time
Notes
a: If you have selected "Brix" or "Solid" (HFSC EPROMs) as a result, only 2 decimal places
are possible.
b. For the report of measured values, the GA42 Printer has space for 7 characters. With 5
places following the decimal point for a concentration value >10, it then prints "9.99999".
However, the result memory contains the correct value.
RA-510M
METTLER TOLEDO
[Method 1]
Variable
Off
:[5]
[Method 1]
nD
:Off On
Form(0-5):[5]
Conc.
:Off On
Form(0-5):[3]
• Refractive index with
maximum 5 decimal places
• Concentration with
maximum 5 decimal places
– Select the desired setting with the
-/Disp. key.
Fix: Fixed record of a result: Series/
sample number, measurement
temperature, refractive index and
concentration are printed out.
nD/Conc. Form: nD is recorded
– With the -/Disp. and ENTER keys, ac-
tivate (On) or deactivate (Off) the de-
sired parameters.
Methods
21

Werbung

Kapitel

Inhaltsverzeichnis
loading