Report of a zero point calibration (GA42 printer)
[Zero Point Calib.]
Model:RA-510M
S/N
:UDB26543
Date :96/05/24 14:19
Temp :
20.00 C
Old
New
1.33301 →
1.33299
Name:Ch. Iserland
Reference table of the refractive index of deonized water
°
Temperature [
C]
15
.............................
20
.............................
25
.............................
6.2.2

2-Point calibration

You usually perform a zero point calibration for normal operation. If you prefer a 2-point
calibration for a defined small region (<0.05), you can perform this with two reference
substances whose refractive indices at the appropriate measurement temperatures are
known. The accuracy will thus be increased for this region; the measured values can be
incorrect outside this region!
The generated line equation is stored, i.e. the characteristic line of the optical sensor is over-
written. To obtain correct values for the whole refractive index region, you must reset this
line equation again (see page 12).
10
**
°C
Temp.
20.54
OK!
**
Temp. 20.04 °C
Name of the refractometer
Instrument identification (see rating plate on the rear panel)
Date and time of the calibration
Measurement temperature of the calibration
The refractive index of water from the last calibration and the one
just performed
User name
n
D
1.33339
1.33299
1.33250
When a stable temperature of 20 °C
± 0.05 is reached, the calibration will last
approx. another minute.
Completion of a successful calibration is
shown in the display and by an audio
signal and is recorded.
If the calibration could not be performed
successfully, Calib. NG! appears.
The refractive index of water is now
shown.
°
Temperature [
C]
30
.................................. 1.33194
35
.................................. 1.33131
40
.................................. 1.33061
n