6. Configuration
Characters,
factory setting
PV
A2
SV
0
EN
PV
A2
SV
1.0
PV
A2
SV
0
PV
A2Ln
SV
nonL
PV
F2
SV
1
PV
F2
SV
00.00
30
Selected object, functions, range
EV2 upper limit value
Setting the upper limit value of EV2
Only possible in the following cases:
When 004 (alarm output, independent of lower/upper limit value alarm),
006 (alarm output, independent of lower/upper limit range alarm) or
012 (alarm output, independent of lower/upper limit value alarm with standby) is
selected with the "Assignment of event output EV2".
Range: Such as with "EV2 limit"
EV2 alarm hysteresis
Setting the EV2 alarm hysteresis
Setting range: 0.1 ... 1,000.0 °C (°F)
DC voltage, current inputs: 1 ... 10,000
Available when an alarm from 001 (alarm output, upper limit value alarm) to
012 (alarm output, independent of lower/upper limit value alarm with standby) is
selected in the "Assignment of event output EV2".
EV2 alarm delay time
Setting the EV2 alarm delay time
When the response time has elapsed, after the input has reached the
alarm output range, the alarm is activated.
Setting range: 0 ... 10,000 seconds
Available when an alarm from 001 (alarm output, upper limit value alarm) to
012 (alarm output, independent of lower/upper limit value alarm with standby) is
selected in the "Assignment of event output EV2".
EV2 switch behaviour alarm
Select the status live/volt-free for the EV2 alarm
Open
nonL
Closed
E8
Available when an alarm from 001 (alarm output, upper limit value alarm) to
012 (alarm output, independent of lower/upper limit value alarm with standby) is
selected in the "Assignment of event output EV2".
TS2 time signal
Setting the program step, at which the time signal output TS2 will be
switched on or off during the program control
Setting range: 1 ... 9
Only available when 015 (time signal output) is selected in "Assignment event output
EV2".
TS2 OFF time
Setting the time signal output TS2 OFF
Setting range: 00:00 to 99:59
Only available when 015 (time signal output) is selected in "Assignment event output
EV2".
WIKA operating instructions models CS6S, CS6H, CS6L