CH1:
1. Press the Cursor button.
2. Select Source in the bottom menu; select ① in the left menu.
3. Select Type in the bottom menu; select Time in the right menu. Two
red dotted lines displayed along the vertical direction of the screen,
which indicating L1 and L2.
4. Select Line in the bottom menu; select L1 in the right menu. The
icon will appear beside the dotted line of L1, represents you can turn
the G knob to adjust the position of L1. Select L2 to adjust the position
of L2. Select L1&L2 to adjust the two cursors together. The cursor
measure window at the top right shows the present time of cursor 1
and cursor 2, the absolute difference and the frequency (see Figure
0-11).
Figure 0-11 Waveform of Time Cursor Measurement
The Cursor Measurement for FFT mode:
Perform the steps below for the Vamp cursor measurement in FFT mode:
1. Press the Cursor button.
2. Select Source in the bottom menu; select ○,M in the left menu.
3. Select Type in the bottom menu; select Vamp in the right menu. Two blue
dotted lines displayed along the horizontal direction of the screen, which
indicating L1 and L2.
4. Select Line in the bottom menu; select L1 in the right menu. The
will appear beside the dotted line of L1, represents you can turn the G
knob to adjust the position of L1. Select L2 to adjust the position of L2.
143
icon