Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance Of Cutting Table; Lubricating The Guide Shafts - struers Axitom Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Maintenance of Cutting Table

Lubricating the Cutting Table

Lubricating the Guide Shafts

X-table Guide Shafts
64
Axitom
Instruction Manual
The stainless steel bands forming the cutting table should be
replaced if they have become worn or damaged. The bands are
available as spare parts.
To allow humidity to escape from the cutting table and chamber, it is
recommended to leave the hood open when the machine is not in
use.
To maintain the optimum performance of Axitom, lubricate the cutting
table at regular intervals (approx every 100 hours).
Check the service info displayed on the screen at start-up to monitor
the actual number of usage hours.
After lubrication of both x- and y-tables, note the date and number of
service hours on the Maintenance Log table.
A bottle with oil for lubricating the guide shafts and slide bearings is
supplied with the Axitom. When all the oil has been used, refill with
eg. Shell TELLUS oil S100.
Lubricate the guide shafts with Oil.
Move the y- table backwards and forwards to distribute the oil
over the whole length of the shafts.
Remove the stainless steel bands located over the guide shafts
(see illustration).
Remove the M6 screws.
Fill the holes with approx. 20ml Oil or until the reservoir is full.
If the oil in the reservoir appears milky;
this indicates that water has entered the oil reservoir.
Call a Struers Service Technician to clean the reservoir.
Re fit the screws.
Important