Warning
The cutting motor is
overheated ! You can
choose to start a cooling
function.
Motor temperature (thermal
load): 100%
Press F1 to start cooling
Press ESC to cancel
Axitom
Instruction Manual
Explanation
#114
Motor temperature has exceeded
its safe working limit. The motor
has been stopped.
Action
Press F1, the motor's built-in fan
will cool the motor. Monitor the
motor's temperature on the display.
Or
Press ESC and wait until the motor
has cooled.