Stop Settings

AutoStop

42
Axitom
Instruction Manual
There are two ways to set the stop position: AutoStop or Stop
position.
When the AutoStop function is selected (this is indicated in the
Cutting display, see below), the machine automatically stops when
the workpiece has been cut through.
For normal cutting, the AutoStop function is recommended.
In the Cutting Menu, use the knob to select Stop position. To set the
Stop position to AUTO, keep turning the knob clockwise until the
arrow above the column has moved to the right of the column.
A small graph appears to the right of the Stop position bar to indicate that
Axitom has detected that cutting has started.
If this graph does not appear then AutoStop will not function.
When the pointer in the STOP column is not on AUTO, Axitom will
only stop when it reaches the preset stop position or the STOP
key is pressed.
Note: AutoStop acts on load changes on the main motor. In some
cases, for example, when using very low feed rate and/or low force,
subtle load changes may not be detected by the machine, preventing
AutoStop from working properly. This may particularly be the case
when cutting soft materials, pipes or workpieces with changing cross-
section.
If AutoStop does not work properly use the Stop position function
instead or set an Additional Cutting Distance to compensate.