When using the blade holders supplied in sterile condition, the assembly and the following
test of the mechanical parts, the locking functions and the blade feed are done only after
inserting the blade holders in the sterile operating area.
— Perform sterilization only in disassembled condition (handle and tubular shaft
separated)!
— The movable handle must be in the unlocked position during sterilization. The jaws
must be open.
The temperature in the autoclave must not exceed 138°C (280°F) and 18 min holding time,
since otherwise the insulation and other plastic parts might be damaged.
Steam sterilization with fractionated pre-vacuum
Minimum requirement
Maximum sterilization temperature
Maximum sterilization time
Drying time
Other times and temperatures can also be used as long as they are within the appropriate
range. The operator must ensure that sterility is maintained after the sterilization process.
Follow the handling and loading recommendations by the manufacturer of the sterilization
device!
WARNING
Risk of serious injury from defective accessories!
A non-functional unit or defective accessories can endanger the patient or user, as well as
affect the intended functions of the unit.
The unit and its accessories must always be kept in flawless working condition. Intact
insulation, cleanliness and potential damage must be visually ascertained.
Units and accessories unfit for further use are to be scrapped.
72
Bipolar Sealing System marSeal 5 with sterile blade holder
3 min at 132°C (270°F)
138°C (280°F)
18 min
according to validation/program
Instructions for Use
Revision 2