KLS Martin marSeal 5 Bedienungsanleitung Seite 64

Bipolares versiegelungssystem mit sterilem klingenträger

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9
5.2
Preparation for Cleaning
The instrument must be cleaned fully disassembled.
Manual and machine cleaning is always required for the instrument.
5.2.1
Disconnecting the Handle from the Tubular Shaft
NOTICE
The red pushbuttons (4) must be fully pushed for disassembly, as otherwise the instrument
could be damaged. Pull the tubular shaft out of the handle in axial direction and with opened
jaw part.
64
Bipolar Sealing System marSeal 5 with sterile blade holder
Press the two red unlock push-
buttons (4) at the top of the
marSeal 5. A slight rotational
movement between handle and
tubular shaft facilitates loosening.
Pull tubular shaft (1) and
handle (7) apart.
Instructions for Use