WARNING
Risk of infection in case of non-sterile handling!
Non-sterile use and improper sterilization can cause severe health risks for the patient!
Sterility of the blade holder can be guaranteed only if the packaging is undamaged and
the seal intact.
Do not use after the expiration date indicated on the packaging.
Insert the sterile-packaged new blade holder only after the instrument has been sterilized
and only in the sterile operating area.
Blade holders are single-use items.
— Remove the blade holder following each application!
— Dispose of used blade holders!
Never reprocess blade holders.
Never reuse blade holders!
WARNING
Risk of burns or tissue damage from operating errors!
Before starting the sealing process, verify that you have grasped the correct structures in
order to prevent inadvertent capture of healthy structures such as ureter, nerve tracts,
intestinal loops, etc.!
Permanent occlusion is possible for vessels up to 7 mm in diameter. Placing two seals side
by side is recommended for large vessels. The integrated blade can then be used to cut
this in the middle.
Ensure that the tissue structures to be sealed are not too thin. For the sealing process,
the amount of tissue must sufficiently fill the space between the electrodes after latching
them!
WARNING
Risk of burns due to high current concentration!
Begin with a low power setting and increase setting if required.
Determine the cause if no or only a poor sealing effect is achieved. Do not increase power
settings up to the maximum without prior examination of the situation!
60
Bipolar Sealing System marSeal 5 with sterile blade holder
Instructions for Use
Revision 2