Instructions for Use
Bipolar Sealing System marSeal 5 with sterile blade holder
WARNING
Risk of electromagnetic interference in case of active implants!
In case of patients with cardiac pacemakers or other active implants, there is always the
danger of interference with or damage to the active implants.
Consult a cardiologist and the manufacturer of the pacemaker or active implant before the
operation!
WARNING
Risk of burns from unintended contact to the patient's tissue!
Unintended activation, e.g. by the foot switch, can lead to burns and electrical shocks if the
electrosurgery instrument is deposited on the patient and thus comes into contact with tissue!
Do not deposit temporarily unused instruments on the patient during the intervals
between applications!
— Instruments should be deposited away from the patient, e.g. on an instrument table!
Never seal unless the jaw part of the instrument is visible! Unintended current activation
or movements of the instrument can lead to patient injury!
Never touch other metallic instruments or objects during the sealing process!
Always make sure the instruments do not come into accidental contact with tissue or
other low-resistance materials (e.g. liquids)!
WARNING
Risk of burns due to hot surfaces!
As a result of heat conduction, the electrodes in the jaw part and the outer surfaces of the
jaw part remain very hot even after the sealing process has been completed, so they can
cause accidental burns.
Avoid touching the jaw part and contact with tissue immediately after the sealing process!
WARNING
Risk of infection in case of non-sterile usage!
Remove and dispose of the blade holder after each usage!
Never reprocess and/or reuse blade holders!
Revision 2
55