3
Safety Notes
WARNING
Non-compliance with the following safety guidelines may result in serious or even
fatal injury!
WARNING
Risk of explosion and fire from highly flammable gases!
Do not use any flammable anesthetic gases or oxidizing gases in connection with the bipolar
sealing system marSeal 5.
Make sure there are no endogenous gases in the application area before beginning the
sealing process.
WARNING
Risk of burns upon contact of the active electrode with metallic parts!
Upon contact of the active electrode with metallic parts, shunts for the HF current or
concentrated leakage current paths can form. These can cause burns.
When activating the current, a safety distance of at least 1 cm must be maintained
between the current-conducting metallic jaw part and the metallic trocar end!
Never touch other metallic instruments or objects during activation!
WARNING
Risk of burns due to high current concentration!
Begin with a low power setting and increase setting if required.
Determine the cause if no or only a poor sealing effect is achieved. Do not increase power
settings up to the maximum without prior examination of the situation!
54
Bipolar Sealing System marSeal 5 with sterile blade holder
Instructions for Use
Revision 2

Werbung

loading