Range and accuracy of measured parameters
Measurement
Humidity
Flow rate
Oxygen concentration*
Respiratory rate
Peripheral blood oxygen
saturation
Pulse rate
* Oxygen measurements are automatically compensated for changes in barometric pressure.
** An RMS accuracy is a statistical calculation of the difference between device measurements and reference measurements.
Approximately two-thirds of the device measurements fell within +/- ARMS of the reference measurements in a controlled study.
Symbol
Displayed Range
Temp
31–37 °C
Flow
2–70 L/min
FiO
21–100%
2
RR
4–70 BPM
SpO
1–100%
2
18–321 beats/min
PR /
Airvo 3 - USER MANUAL | A - 45
Accuracy
Not specified
± (1 + 5% of reading) L/min
Lower of:
± 4%, or
± (2.5% + 2.5% of reading)
- excluding rounding to 21% and 100%, as appropriate
- provided "Oxygen concentration" setting is correct
RMS error of <3 BPM**
See Nonin pulse oximetry specifications below.
See Nonin pulse oximetry specifications below.