Alarm condition
Priority
High FiO
(HPO)
Low
2
Pulse oximetry alarms
Pulse Ox Disconnected
Med.
Pulse Ox Communication
Med.
Failure
No Pulse Ox Sensor
Med.
Connected
Pulse Ox Sensor Off
Med.
Patient
No SpO
Reading
Med.
2
No Pulse Rate Reading
Med.
Pulse Ox Not Recognized
Med.
Physiological alarms
Low SpO
High
2
High SpO
Med.
2
Disinfection alarms
Disinfection Failed to
Med.
Hold Temperature
Over Temperature
Med
Detected
Delay
Meaning
<5 s
The FiO₂ target is above the selected High Oxygen Alarm threshold for its
intended clinical environment (range 30–95% or Off, default: 95%, see Airvo 3
Technical Manual). Check FiO
FiO
to the normal range when it is appropriate to do so.
2
<1 s
The pulse oximetry accessories have become disconnected. Reconnect pulse
oximetry accessories.
<10 s
The Airvo 3 is unable to communicate with the pulse oximeter. Check that the
USB connector cable, sensor adapter cable and sensor cables are all properly
connected. Replace the sensor cable, adapter cable then USB connector cable in
turn if the problem persists.
<1 s
A pulse oximetry sensor cable was not detected or is inoperable. Check that the
sensor cable is properly connected to the USB connector cable or replace the
sensor cable if necessary.
<1 s
The pulse oximeter is no longer receiving SpO
Check that the sensor is properly attached to a suitable measurement site on
the patient.
<16 s
The pulse oximeter is not sending valid SpO
Check the sensor, cable and USB interface. Try replacing each component in turn
until the problem is resolved.
<16 s
The pulse oximeter is not sending valid Pulse Rate measurements.
Check the sensor, cable and USB interface. Try replacing each component in turn
until the problem is resolved.
<10 s
The selected pulse oximeter has not been recognised. Please remove or change
oximeter.
<15 s
Check your patient. The SpO
alarm threshold. Check that the alarm setting is appropriate for your patient
(Range: 1–98%, default 85%, see Airvo 3 Technical Manual).
<15 s
Check your patient. The SpO
threshold. Check that the alarm setting is appropriate for your patient (Range:
2–99% or Off, default Off, see Airvo 3 Technical Manual).
<3 min
The Airvo 3 cannot heat up to the required disinfection temperature. Check:
• the disinfection tube blue connector is connected to the top of the outlet
elbow,
• the disinfection tube red end is connected to the left hand chamber port,
• the disinfection filter is connected to the right hand chamber port,
Then restart the device. If the problem is not resolved, replace the disinfection
tube and outlet elbow in turn. If the issue persists, contact your service
representative.
<5 s
The Airvo 3 detected higher than expected temperatures during the
disinfection cycle. Check:
• the disinfection tube blue connector is connected to the top of the outlet
elbow,
• the disinfection tube red end is connected to the left hand chamber port,
• the disinfection filter is connected to the right hand chamber port,
Then restart the device. If the problem is not resolved, replace the disinfection
tube and outlet elbow in turn. If the issue persists, contact your service
representative.
Airvo 3 - USER MANUAL | A - 31
is appropriate for the patient's condition. Reduce
2
measurements from the patient.
2
measurements.
2
measurement has decreased below the Low SpO
2
measurement has exceeded the High SpO
2
English
2
alarm
2