A - 26 | Airvo 3 - USER MANUAL
6.1
Patient data
The values displayed in the patient data screen are described below. Measurements
that are not available are shown as "--". Measurements may not be available when the
Airvo 3 is in standby mode or the device has not collected enough data for a reliable
measurement.
Label
Unit
Flow
L/min
RR
BPM
Humidity
°C
FiO
%
2
SpO
/FiO
2
2
ROX
SpO
%
2
PR
BPM
6.2 Live value graphs
Airvo 3 live value trends graphs show the last 30 seconds of data.
Label
Unit
Flow
L/min
6.3 Long term graphs
Airvo 3 Data and Graphs show measurements plotted against time for up to 24 hours. New measurements are added to the right side of
the graph. Prior data will scroll to the left as new measurements are added. Gaps will appear in the plotted data if therapy is stopped or
measurements are missing due to poor signal quality.
The graphs available are described in the table below.
Label
Unit
Target flow
L/min
RR
BPM
FiO
%
2
SpO
/FiO
2
2
ROX
SpO
%
2
PR
BPM
Description
The current flow rate of breathing gases supplied to the patient
The patient's respiratory rate (breaths per minute), averaged over the last 90 seconds
The current humidity of breathing gases supplied to the patient
The current fraction of oxygen in breathing gases supplied to the patient
Ratio of SpO
and FiO
2
2
SpO
divided by FiO
and respiratory rate
2
2
Peripheral blood oxygen saturation measured by pulse oximeter
Pulse rate measured by pulse oximeter (beats per minute)
Description
The current flow rate of breathing gases supplied to the patient
Description
The target flow rate of breathing gases supplied to the patient
The patient's respiratory rate (breaths per minute), averaged over the last 90 seconds
The fraction of oxygen in breathing gases supplied to the patient
Ratio of SpO
and FiO
2
2
SpO
divided by FiO
and respiratory rate
2
2
Peripheral blood oxygen saturation measured by pulse oximeter
Pulse rate measured by pulse oximeter (beats per minute)