A - 34 | Airvo 3 - USER MANUAL
8.1.2
Disinfect
Perform disinfection only after all cleaning steps are complete
Equipment
• Disinfectant wipes
• Clean, disposable, lint-free cloths
• Protective gloves
Disinfection Instructions
1. Use pre-soaked disinfecting wipes, to thoroughly wipe all outside surfaces of the device (including the Outlet Elbow).
2. Ensure that these surfaces remain visibly wet as directed by the manufacturer of the disinfecting wipes. Use additional wipes as needed
to ensure that these surfaces remain wet for the required length of time.
Rinse
3. Dampen a clean cloth with tap water.
4. Thoroughly wipe all outside surfaces of the device with the damp cloth to remove any disinfectant residue.
Dry
5. Thoroughly wipe all outside surfaces of the device with a dry cloth until it is visibly dry.
6. Allow to air dry until completely dry.
Warnings
Other cleaning agents may be used if they are: non-abrasive, non-toxic, and non-corrosive. Do not use any cleaning agents that are
not compatible with polycarbonate plastic.
Cleaning agents that are not suitable for use with the Airvo 3 include: ammonia, ammonium hydroxide, caustic soda, iodine,
methanol, methylated spirits, turpentine, and alkaline bleaches such as sodium hypochlorite. The use of any of these products will
damage the Airvo 3.
Turn off and disconnect the Airvo 3 from the wall power supply before cleaning to reduce the risk of electric shock.
Do not submerge the device in liquid of any kind.
Do not spray liquid directly onto the device.
These instructions have been validated by the manufacturer of the medical device as being capable of preparing a medical device for
reuse. It remains the responsibility of the processor to ensure that the processing achieves the desired results, by using the correct
equipment, materials, and personnel in the processing facility. This requires routine monitoring of the process.
8.2
Outlet elbow reprocessing
The Outlet Elbow requires cleaning and high-level disinfection. The Outlet Elbow can be reprocessed in two diff erent ways:
– Disinfection Kit 900PT600 (see instructions in 900PT600)
– Washer-Disinfector (follow below instructions)
The outlet elbow can be removed from the Airvo 3 for reprocessing by your central sterile services or reprocessing department.
Reprocessing the outlet elbow must be performed in a washer-disinfector that complies with and is maintained, checked and validated to
ANSI/AAMI ST15883-1:2009 (USA) and ISO 15883-1:2006 (outside USA).
Disassembly
Remove the outlet elbow from the Airvo 3. Firmly grab the rubber port-seal on the
outlet elbow, push down on the grip lines with your thumb and pull the outlet elbow
towards the front of the Airvo 3.