A - 28 | Airvo 3 - USER MANUAL
7.4
Viewing alarm details
Alarms are displayed with suggestions and action buttons
for managing the alarm information:
• Tap the Audio Pause button to silence the alarm for
120 seconds. The Audio Pause button will change to
when audible alarms are silenced.
• Use
to scroll through multiple suggestions.
Some alarms have only one suggested resolution.
• Tap Hide suggestions to collapse the alarm
information to the Message bar. Restore
suggestions by tapping the alarm condition
on the Message bar.
The alarm condition and action button are displayed
on the Message bar when alarm information is
collapsed.
The Message bar cycles through each alarm condition
when multiple alarms are active. Tapping the Message
bar displays a list of active alarm conditions, from
highest priority to lowest priority and they are ordered
from when they occurred.
Alarm signals always indicate the highest-priority
active alarm condition.
7.5
Checking the alarm system
To test the alarm system:
1. In standby mode, disconnect the breathing tube
then press "Start".
2. Verify that the "Check tube" visual alarm appears
on screen.
3. Verify that the signal light fl ashes yellow.
4. Verify that the auditory alarm signal can be
heard.
Do not use the Airvo 3 if it fails this test. Contact
your Fisher & Paykel Healthcare representative.
7.6
Airvo 3 alarms
All the alarms you may encounter while using the
Airvo 3 are listed below with common causes, resolutions and any delays inherent in determining alarm conditions. These priorities have
been allocated for an intended operator's position of 2 meters from the device. The Airvo 3 also uses an internal-priority ranking system.
If multiple alarm conditions occur simultaneously, the device will display the highest-priority alarm.
Alarm condition
Priority
Faults
Device Fault
High
[Fault X.X.X]
Device Fault
Med.
[Fault X.X.X]
Power alarms
Power Out
High
Alarm condition
Action buttons
Alarm condition
Action buttons
Show suggestions
Hide suggestions
Reset & dismiss
Reset & dismiss
Number of
active alarms
Delay
Meaning
-
A technical fault has occurred that has caused therapy to be stopped. Restart
the device to resolve the fault. If the issue persists, contact your service
representative.
-
A technical fault has occurred, and the device is able to continue supplying
limited therapy. Restart the device to resolve the fault. If the issue persists,
contact your service representative.
<5 s
The Airvo 3 has been disconnected from the wall power supply and the internal
battery is depleted. The auditory alarm will sound once every 10 seconds for
120 seconds and the signal light above the touch screen will fl ash. The touch
screen is off during the Power Out alarm. The Airvo 3 will shut down after
signaling the Power Out alarm but will restart automatically if power is restored
before it shuts down.
Hide
suggestions
Suggestions/
action
Mute
Audio pause
Alarm co
Mute ala
Mute alarm
Unmute
Unmute alarm
Suggestions/
action
Active alarms
Audio pause
Hide
Sugg