Review disinfection state
The Airvo 3 will ask you if it will be used on:
the same patient who last used the device (tap Same Patient)
OR
a new patient (tap New Patient).
For a new patient, check that:
1. The outlet elbow has been cleaned and disinfected.
2. A new tube and chamber have been installed.
Review disinfection state
(if the disinfection method is set to Disinfection kit only)
For a new patient, check that:
1. The outlet elbow has been cleaned and disinfected.
The Airvo 3 will indicate the outcome of the last disinfection cycle:
Green: The previous disinfection cycle was completed successfully.
Orange: A successful disinfection cycle has not been performed.
Please run a successful disinfection cycle before use on a new
patient.
Red: The previous disinfection cycle failed to complete. Please run
a successful disinfection cycle before use on a patient.
The number of successful disinfection cycles completed by the Airvo 3 is
displayed in the lower left hand corner under 'Disinfection count'.
2. A new tube and chamber have been installed.
English
Airvo 3 - USER MANUAL | A - 17