Teach function of the limit value switches
S
Zero settings
S
Any selected conversions to another unit (measurement is completed in N again)
S
Minimum and maximum values, as well as all of the statistical information (peak-
peak) will be lost.
S
Settings for digital inputs and outputs
S
Deactivation of the nominal overload warning is activated
3. Restore factory reset
The sensor does not restart. In addition to the parameters specified in the application
reset, the entries in the fields "Application Tag", "Function Tag", and "Location Tag" are
reset.
Any linearization that may have been entered in the sensor (calibration certificate) is lost.
4. Back to box
All of the parameters are lost. Any force overshoots will continue to be saved. The sensor
will return to the delivery condition. Any linearization that may have been entered in the
sensor (calibration certificate) is lost.
System commands
The IO-Link standard defines some of the "System Commands". Further application-
specific commands are added to the standard commands by the electronics.
Index
Sub­
(hex)
index
(hex)
0x0002
0
A command is triggered by writing the assigned code on the "System Command" vari­
able. The electronics supports the following commands:
Code
Function
(decimal)
65
Teach switching point limit value
switch 1
66
Teach switching point limit value
switch 2
128
Device Reset
129
Application Reset
130
Restore factory settings
U9C
ELECTRICAL CONNECTION
Permissions
Write Only
Data
Data
Type
Size
bytes
UINT8
1
See section
7.2.6 Limit value switch
7.2.6 Limit value switch
Reset functions
Reset functions
Reset functions
Description
System Command
51

Werbung