Herunterladen Inhalt Inhalt Diese Seite drucken

Alarms (Io-Link Events) - HBK U9C Montageanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Index
Sub­
(hex)
index
(hex)
0x0D49 0x00
0x0D4A 0x00
0x0D4B 0x00
0x0D4C 0x00
0x0D4D 0x00
0x0D4E 0x00
0x0D4F 0x00
Index
Sub­
(hex)
index
(hex)
0x0002
0x00

8.3.11 Alarms (IO-LINK events)

The electronics monitor the sensor and compare the mechanical and thermal stresses on
a continuous basis against the limit values of the force sensor. In the event of thermal
monitoring, these are also compared to the limit values of the electronic components.
The electronics uses a very high sample rate to evaluate the mechanical stress. Even very
short maximum forces are recorded and a message will be issued if the limit values are
exceeded. As the output of measurement values via the digital data interface/IO-Link
connection is completed at a lower sample rate, it may be the case that you cannot find a
force value, that was registered as a force overshoot, in the measurement data.
U9C
ELECTRICAL CONNECTION
Permissions
Data
Type
ReadOnly
UIntegerT
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
Permiss­
Data
ions
Type
Write
UINT
Data
Description
Size
bytes
8
statistic1 n
4
statistic1
last_value
4
statistic1
min_value
4
statistic1
max_value
4
statistic1
pp_value
4
statistic1
mean_value
4
statistic1
stddev_value
Data
Description Value
Size
bytes
1
Statistics
reset
Description
Number of
measurement
values since the
last reset
The current mea­
surement value as
a sample that is
used as an entry
for the statistical
calculations.
Minimum value
Maximum value
Peak-to-peak
value
Mean value
Standard
deviation
Description
(decimal)
209
Restart
recording of
statistical
values
43

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für HBK U9C

Inhaltsverzeichnis