8.3.11 Alarms (IO-LINK events)
The electronics monitor the sensor and compare the mechanical and thermal stresses on
a continuous basis against the limit values of the force sensor. In the event of thermal
monitoring, these are also compared to the limit values of the electronic components.
The electronics uses a very high sample rate to evaluate the mechanical stress. Even very
short maximum forces are recorded and a message will be issued if the limit values are
exceeded. As the output of measurement values via the digital data interface/IO-Link
connection is completed at a lower sample rate, it may be the case that you cannot find a
force value, that was registered as a force overshoot, in the measurement data.
U9C
ELECTRICAL CONNECTION
Permissions
Data
Type
ReadOnly
UIntegerT
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
ReadOnly
Float32T
Permiss
Data
ions
Type
Write
UINT
Data
Description
Size
bytes
8
statistic1 n
4
statistic1
last_value
4
statistic1
min_value
4
statistic1
max_value
4
statistic1
pp_value
4
statistic1
mean_value
4
statistic1
stddev_value
Data
Description Value
Size
bytes
1
Statistics
reset
Description
Number of
measurement
values since the
last reset
The current mea
surement value as
a sample that is
used as an entry
for the statistical
calculations.
Minimum value
Maximum value
Peak-to-peak
value
Mean value
Standard
deviation
Description
(decimal)
209
Restart
recording of
statistical
values
43