If you are working with a square approximation, please set R to zero. If you are using a
linear approximation, please set R and S to zero. The calibration certificate must contain
tared values, i.e., the cubic function must not contain any constants.
In the "Adjustment" menu, the name of the calibration facility, date, calibration certificate
designation, and expiry date can be saved in the electronics for documentation purposes
if a calibration certificate is available. (see above)
Index
Sub­
(hex)
index
(hex)
0x0C2A 0x02
0x0C2A 0x03
0x0C2A 0x04
0x0C2B 0x02
0x0C2B 0x03
0x0C2B 0x04
Measurement output using a different unit
Use the "Unit Conversion" point to select a different unit than N. The numeric value that is
then sent to the downstream electronics is the same as the one displayed in the software
of your IO-Link master.
You can now select the unit under "Process data switch". If you select kN, MN the
conversion is completed automatically; if you select a different unit you will be asked to
enter a "Unit Conversion Factor".
Example: Output should be in kg. Select kg as the unit. At your site the Gravitational
acceleration is 9.806 m/s
The calculation is then completed: Output in kg = measurement value in
2
N x 0.101979 s
/m
If you have to complete a conversion, that requires a constant factor, then you can enter it
under "Zero Offset User".
34
Permissions
Data
Type
ReadWrite
Float32T
ReadWrite
Float32T
ReadWrite
Float32T
ReadWrite
Float32T
ReadWrite
Float32T
ReadWrite
Float32T
2
. The scale factor 1/9.806 m/s
Data
Description
Size
bytes
4
Adjustment
coeffs T
compr.
4
Adjustment
coeffs S
compr.
4
Adjustment
coeffs R
compr.
4
Adjustment
coeffs T tens.
4
Adjustment
coeffs S tens.
4
Adjustment
coeffs T tens.
2
= 0.101979 s
Description
Linear portion for
the pressure
range
Square portion
for the pressure
range
Cubic portion for
the pressure
range
Linear portion for
the tensile range
Square portion
for the tensile
range
Cubic portion for
the tensile range
2
/m
ELECTRICAL CONNECTION