Important
Please remember that the linearization is only effective if "Linearization" is NOT set to
"disabled"
Index
Permissions
(hex)
0x0C44 ReadWrite
0x0C45 ReadWrite
0x0C46 ReadWrite
0x0C47 ReadWrite
0x0C26 ReadWrite
Linearization via supporting points
u
Select "Stepwise linear adjustment" and the "Adjustment supporting points" menu is
displayed.
u
Enter the number of supporting points, this number can be between 2 and 21. Please
note that the zero point is a supporting point. If you want to enter a straight line, you
must select two supporting points.
u
Enter the force generated by the calibration machine for the relevant level in the field
"Ext Cal X".
u
Copy the measurement result of the calibration from the calibration certificate and
enter the numerical values as "Ext Cal Y".
u
It is important to start with the most negative force as that is the highest tensile force.
On pure compressive force sensors , 0 N is defined as the "highest tensile force".
32
Data
Data
Type
Size
bytes
StringT
32
StringT
32
StringT
32
StringT
32
UIntegerT 1
Description
Description
Calibration
Calibration date
date
Calibration
Calibration
authority
laboratory
Calibration cer­
Number of the cali­
tificate ID
bration certificate
Calibration
Date on which a new
expiration date
calibration is
required
Linearization
Selection of the type
of linearization:
0: no linearization
applied
1: linearization via
supporting points
2: linearization via
cubic function
ELECTRICAL CONNECTION