CONDITIONS ON SITE

U9C series force transducers are made of rustless materials. It is nevertheless important
to protect the transducers from weather conditions such as rain, snow, ice and salt water.
6.1

Ambient temperature

The effects of temperature on the zero signal and on sensitivity are compensated.
To obtain optimum measurement results, comply with the nominal (rated) temperature
range. The compensation of the temperature effect on the zero point is implemented with
great care, but temperature gradients can still have a negative effect on the stability of
zero. Constant or very slowly changing temperatures are therefore best. A radiation
shield and thermal insulation on all sides provide marked improvement, However, they
must not be allowed to set up a force shunt, i.e. slight movement of the force transducer
must not be prevented.
6.2

Moisture and corrosion protection

The force transducers are hermetically encapsulated and are therefore very insensitive to
moisture. The transducers achieve protection class IP67.
Despite the careful encapsulation, it makes sense to protect the transducers against
permanent exposure to moisture.
The force transducer must be protected against chemicals that could attack the steel.
With stainless steel force transducers, note that acids and all materials which release
ions will in general also attack stainless steels and their welded seams. Should there be
any corrosion, this could cause the force transducer to fail. In this case, appropriate
protective measures must be provided.
The housing of the inline amplifiers is made of aluminum and, like the sensor, meets the
requirements of IP67 protection. We recommend protecting the amplifier housing from
sustained weather effects.
6.3

Deposits

Dust, dirt and other foreign matter must not be allowed to accumulate sufficiently to
conduct any of the measuring force around the transducer, thus invalidating the
measured value (Force shunt). Also remember to lay the connection cable so that no
force shunts are generated at the lower nominal (rated) forces (<1 kN).