8.5.1.3 Testing the Water Meter
Entry into programming level.
(Also refer to chapter
Go to Manual Functions
Selection of the "Solenoid Valve" function (
Water meter function
is used to open the solenoid valve and to display the current water throughput.
is used to close the solenoid valve and to display the after-run. This may not
exceed 250 ml! If necessary, replace the solenoid valve.
Exit programming mode again.
8.5.2
Bleeding the Metering Pump / Manual Functions
Bleed the metering pump as follows:
Undo the swivel nut at the admission point, remove the flow monitor
(see chapter 5.1, Fig. 5.1, Item 15) and place it in a suitable collection receptacle.
Go to the programming level.
(Also refer to chapter
Go to Manual Functions
Selection of the "Pump" function (
is used to start the metering pump.
is used to stop the metering pump.
Exit programming mode again.
The solenoid valve
test.
417102242_DG_3
8.2
"Programming Structure" and
Fig. 8.58
8.2
"Programming Structure" and
)
Fig. 8.59
Bleeding the metering pump / Manual functions
can also be switched on and off similar to the metering pump
- 85 -
8.4.1
"Operation Handling").
)
Water meter test
8.4.1
"Operation Handling").
Metering pump
Rev.: 02-05.2014