8.4.5.1 Setting concentration 1
Check the concentration setting as follows and correct it, if necessary:
Unfasten the flow monitor (see chapter 5.1, Fig. 5.1, Item. 15) and seal the opening
using the provided sealing plug.
Insert the flow monitor into the provided measuring cylinder and place a bucket under
the swivel spout.
Activate the "Calibrate" menu, set the required tap quantity and start the tapping
process by pressing
(Also refer to chapter
"CAL 1" = Calibration
of concentration 1
Check the quantity of disinfectant in the measuring cylinder.
Measure the quantity of water tapped. This must correspond to the target quantity less
the quantity of product.
By pressing
NOTE
calibration function.
Calculate the concentration using the formula (see above). If your result is different from
the required concentration (taking the safety factor into account), correct the DLA steps
in the display below in percentage terms relative to the DLA value shown.
Save the new value
If concentration 2 should also be set, continue with
exited by pressing
8.4.5.2 Setting concentration 2
Concentration 2 is set in the same way as concentration 1.
(see chapter 8.4.5.1, chapter
Following completion of calibration, remove the sealing plug from the opening and
screw the flow monitor back in.
417102242_DG_3
.
8.2
"Programming Structure" and
Fig. 8.41
or
, the user can also switch on directly to the concentration 2
Fig. 8.42
Setting concentration 1 (2)
.
8.2
"Programming Structure" and
Setting concentration 1
Where necessary, correct the
- 79 -
8.4.1
"Operation Handling").
Setting the required
tapping quantity
; otherwise, the calibration can be
8.4.1
"Operation Handling").
Rev.: 02-05.2014