Herunterladen Inhalt Inhalt Diese Seite drucken

Transducer Connection Side Leds; Overflow / Underflow Indication - HBM espressoDAQ DQ401 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für espressoDAQ DQ401:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Meaning of the LEDs

8.3.2 Transducer connection side LEDs

The module run-up phase begins after the supply voltage is connected.
This is indicated by all the LEDs on the transducer side of the module succes-
sively lighting up.
Sequence:
1. All the red LEDs light up
2. All the orange LEDs light up at one-second intervals
3 One second later: all the green LEDs light up
The LEDs then indicate the current status at the transducer connection.

8.4 Overflow / Underflow indication

for module DQ401:
Voltage:
"10 V
Current:
4 − 20 mA
"25 mA
for module DQ809:
Temperature:
"100 mV
o
C
"4096
A2888−1.1 en/de
LED A (channels 1, 2, 3, 4)
Overflow: +12 V (120%)
Underflow: −12 V (−120%)
if a transducer is not plugged in, a value of +12 V is shown
and indicated by a red LED
LED B (channels 1, 2, 3, 4)
Overflow: +22 mA (110%)
Underflow: +2 mA (−110%)
Overflow: +32.5 mA (130%)
Underflow: −32.5 mA (−130%)
LED A (channels 1, 3, 5, 7)
Overflow: +205 mV
Underflow: −205 mV
Overflow: + 8400
Underflow: − 8400
LED B (channels 2, 4, 6, 8))
o
C
o
C