Transducder connection

7.5.1 Thermoadapters

To operate the DQ809 with thermocouples, the 1-KAB419 thermoadapter is
connected to the relevant transducer connector socket (see page 39).
Two thermocouples can be connected to one thermoadapter.
The identifying letters A and B on the thermoadapter make the distinction bet-
ween the two channels. Letter A always marks the odd channel number and B
the even channel number in the catman
The cold junction is located in the thermoadapter.
The following thermocouple types can be connected:
J , K , T , S , E , B , N , R , C
Table of all types with materials, temperature range, color
Type
J
K
T
S
E
B
N
R
C
Relevant color coding or type coding can be attached to the offset area below
the thermocouple sockets.
A different thermocouple type can be defined for each channel!
A2888−1.1 en/de
Material
Fe-CuNi
NiCr-NiAl
Cu-CuNi
Pt10Rh-Pt
NiCr-CuNi
Pt30Rh-Pt6RH
NiCrSi-NiSr
Pt10Rh-Pt
Wo5Rh-Wo26Rh
Starter channel card.
Temperature range
(continuous operation)
+20 − +700
0 − +1100
−185 − +300
0 − +1550
0 − +800
+100 − +1600
0 − +1100
0 − +1600
0 − 2320
41
Color
black
green
brown
orange
violet
gray
pink
orange