Operation
➜
Confirm or discard all changes with
Process parameters
Process parameters
➜
Touch the
➜
Tap the button of the desired parameter in the menu and adjust the value using the
on-screen keyboard, see also "User input" on page 104.
Parameters
Heating tape
Heating tape
automatic
Evacuation after
draining the
collector
Warm-up time,
filling in
sensor mode
Educt empty:
additional evap-
oration time
Max. filling:
additional evap-
oration time
Accept
button in the footer to open the menu.
Function
On-period of the heating tape in operation.
Range: 10 – 100% in increments of ten.
(see "Heating tape (optional)" on page 96)
Automatically switch on heating tape yes/no.
Range: on / off.
Evacuation period of the bypass line after draining the
collector vessel.
Range: 1 – 9999 s.
Warm-up time when filling the evaporation flask in sensor
mode.
Range: 1 – 9999 s.
When the educt container runs empty: Additional evapo-
ration time for evaporation of the remaining educt in the
evaporation flask.
Range: 1 – 9999 s.
Maximum filling rate: Selectable when a specific number of
filling operations is specified in the start window:
In the example: five fillings, then optionally an additional
evaporation time runs. This allows a desired concentration
to be achieved!
max. filling: Add. evaporation
The parameters
time
max. drain distillate: additional
and
evaporation time
be activated at the same time!
132
Cancel
or
.
are mutually exclusive and cannot