Commissioning
Settings
Open the
also "Main menu" on page 110).
Follow the instructions on the display.
OR
Calibration on high-polar solvents and mixtures:
Make sure the sensor is in the flask.
Fill the evaporation flask with at least 500 ml of the product.
Slide the sensor into the stopped flask until the sensor head is just above the fluid
level or – for example, in the case of mixtures with water – until the sensor head
touches the fluid level slightly.
Start the calibration process in the device software:
Open the main menu using the
the start window:
Settings
Open the
also "Main menu" on page 110).
Follow the instructions on the display.
PRACTICAL TIP
When processing solvents with a tendency toward boiling retar-
dation (e.g. ethanol), it is recommended to calibrate the sensor
slightly less sensitively.
This prevents the boiling retardation from being detected as a fluid
and the fluid level in the rotary flask from decreasing excessively
before adding new medium.
The sensor should only respond to liquids, never to boiling retarda-
tion or foam!
Calibrate filling level
menu and select
Main menu
Calibrate filling level
menu and select