Commissioning
The quick-action chuck can be overturned by excessive force
applied on opening! In this case, the quick-action chuck can no
longer be closed normally and a latching noise can be heard on
turning it in the closing direction.
Remedial action: Turn the quick-action chuck again in the opening
direction, beyond the latching point.
The quick-action chuck can then be operated normally again.
Power supply
The device has an appliance outlet with locking system for the power supply. A suitable
three-pole power supply cord is included in the scope of delivery.
DANGER
Electric shock, damage to property, loss of production
➜
Power cords without a locking system can become loose during
operation due to vibrations. In the event of an uncontrolled shut-
down of the stirring tool, there is a risk of production downtime and
or property damage, e.g. due to uncontrolled restart!
➜
Connect the power supply cord
➜
Before connecting the power supply cord, make sure that the main switch of the
device is in the OFF position (see following section "Switching the device on/off" on
page 66).
➜
Connect the cable coupling of the cord to the IEC appliance inlet on the back of the
device. Make sure that the locking system engages correctly.
➜
Connect the power supply cord to a properly secured mains socket-outlet.
The device may only be supplied with the required line voltage
via a grounded power socket.
For reasons of operational safety, only use the supplied power
supply cord with locking system!
65