Operation

Hei-TORQUE Ultimate 400

Stop the rotational movement.
To switch between gear stage I and II, press the [
the control panel:
The set gear stage is shown on the display.
When working through a profile/favorite the gear stage is selected
automatically.

Starting/stopping rotation

WARNING
Risk of injury due to glass break and leaking medium
Risk of injury on rotating parts
Property damage, loss of production
Select the diameter of the laboratory vessel so that the stirrer
tool turns inside the vessel without touching it.
After clamping the stirrer tool and before the process starts,
check the stirrer tool and the chuck for smooth running.
Never start up stirring tools or chucks that have an unbalance,
and replace them immediately.
Never grip turning parts.
Always wear close-fitting clothing when working with the device.
Do not wear any accessories such as ties, scarves, shawls,
pieces of jewelry, etc. when working with the device.
When working with the device, tie back long hair or wear a cap
or hair net.
Adjust the start-up behavior of the device to the specific
properties of the sample to be processed to avoid overflowing/
ejection of the sample or overloading of the stirrer tool (safety
settings!).
Follow the specific safety instructions for the substances to be
processed (refer to the safety data sheets if applicable).
Always wear the required individual personal protective
equipment (PPE).